title = "3 Dimensional power receiver for in vivo robotic capsules",
abstract = "A 3D wireless power receiver is introduced for in vivo robotic capsules. It is designed to retrieve hundreds mW of power regardless of its orientation relative to the magnetic field generated from the outside of the body. Design issues such as the antenna structure and the rectification topology are discussed considering physical space limitation. Various design parameters are implemented and experimented. The selected design is implemented and tested for the power variations on the rotational positions relative to the transmitter magnetic field. The presented receiver delivers more than 300 mW regardless of its orientation.",
keywords = "3D wireless power receiver, In vivo robotic capsule, Rectifying, Regulating",
author = "Kim, \{Jong Dae\} and Munho Ryu and Jongwon Kim and Chin, \{Hyun Uk\}",
note = "Publisher Copyright: {\textcopyright} International Federation for Medical and Biological Engineering 2007.; 10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 ; Conference date: 27-08-2006 Through 01-09-2006",
year = "2007",
doi = "10.1007/978-3-540-36841-0\_184",
language = "English",