Abstract
This article describes a preliminary investigation and prototype fabrication of a novel potential sensing method to continuously monitor vascular stent fractures. A potential measurement system consisting of Wheatstone bridge circuit and signal conditioning circuit was designed for the cardiovascular stent durability and fatigue test. Each end of a bare and polyurethane-covered Nitinol vascular stent was electrically connected to the potential measurement system and then immersed either in simulated body fluid (SBF) media or distilled water at 36.4±1°C. When the stent experienced fracture (i.e., a cut), its electrical potential decreased with an increase in electrical resistance. This method successfully measured fractures in the stent regardless of location. Furthermore, the number of cycles at the onset of stent fracture was accurately detected and continuously monitored using this technique. Thus, the present fracture detection method, which to our knowledge is the first ever report to use electrical potential measurement for stent durability test, gives a fast, real-time, accurate and efficient detection of fractures in stent during in vitro fatigue and durability test.
| Original language | English |
|---|---|
| Pages (from-to) | 213-222 |
| Number of pages | 10 |
| Journal | Bio-Medical Materials and Engineering |
| Volume | 21 |
| Issue number | 4 |
| DOIs | |
| State | Published - 2011 |
Keywords
- Electrical potential
- fatigue and durability test
- Nitinol
- stent fracture
- vascular stent
- Wheatstone bridge circuit
Quacquarelli Symonds(QS) Subject Topics
- Materials Science
- Biological Sciences
Fingerprint
Dive into the research topics of 'A novel electrical potential sensing method for in vitro stent fracture monitoring and detection'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver