Abstract
Purpose: Four key bench-top tests, including trackability, conformability, wall-apposition, and bending stiffness, were performed to understand the mechanical characteristics in 3 different types of stents applicable for treatment of intracranial atherosclerotic stenosis: Balloon-expandable D+Storm, Pro-Kinetic Energy, and self-expandable Wingspan stents. Materials and Methods: Trackability was assessed by measuring the tracking forces of each stent with its delivery systems. Conformability and wall apposition were quantified and ana-lyzed using curved vessel models. A 3-point bending test was employed to evaluate bending stiffness. Results: D+Storm showed the lowest tracking forces while the conformability of the Wingspan stent was superior to that of the tested stents. Pro-Kinetic Energy and D+Storm had better wall apposition in curved vessels than the Wingspan stent. Bending stiffness of the Wingspan stent was notably lower, whereas no significant differences were found between D+Storm and Energy. Pro-Kinetic Energy and D+Storm not only indicated lower gap ratios between the struts and the vessel wall but also maintained good wall apposition even in the curved model. Conclusion: These bench-top measurements may provide clinicians with useful information in regard to selecting suitable stents for treatment of intracranial atherosclerotic stenosis.
| Original language | English |
|---|---|
| Pages (from-to) | 117-125 |
| Number of pages | 9 |
| Journal | Neurointervention |
| Volume | 15 |
| Issue number | 3 |
| DOIs | |
| State | Published - 2020.11 |
Keywords
- Atherosclerosis
- Intracranial arterial disease
- Stent
Quacquarelli Symonds(QS) Subject Topics
- Medicine
Fingerprint
Dive into the research topics of 'Bench-Top Comparison of Three Different Types of Stents Used for Treatment of Intracranial Atherosclerotic Stenosis'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver