Clinical usefulness of the serum anion gap

  • Sik Lee
  • , Kyung Pyo Kang
  • , Sung Kyew Kang*
  • *Corresponding author for this work

Research output: Contribution to journalJournal articlepeer-review

Abstract

The anion gap in the serum is useful in the interpretation of acid-base disorders and in the diagnosis of other conditions. In the early 1980s, ion-selective electrodes for specific ionic species were introduced for the measurement of serum electrolytes. This new method has caused a shift of the anion gap from 12±4 mEq/L down 6±3 mEq/L. It is worthy for clinicians to understand the range of normal anion gap and the measuring methods for serum sodium and chloride in the laboratories that support their practice. While an increase in the anion gap is almost always caused by retained unmeasured anions, a decrease in the anion gap can be generated by multiple mechanisms.

Original languageEnglish
Pages (from-to)44-46
Number of pages3
JournalElectrolyte and Blood Pressure
Volume4
Issue number1
DOIs
StatePublished - 2006

Keywords

  • Anion gap
  • Ion-selective electrode