Comparison of Iodine Density Measurement among Dual-Energy Computed Tomography Scanners from 3 Vendors

  • Hyungjin Kim
  • , Jin Mo Goo*
  • , Chang Kyung Kang
  • , Kum Ju Chae
  • , Chang Min Park
  • *Corresponding author for this work
  • Seoul National University

Research output: Contribution to journalJournal articlepeer-review

Abstract

Objectives The aims of this study were to analyze the effect of dual-energy computed tomography (DECT) scanners and fluid characteristics on iodine quantification and to calculate the measurement variability range induced by those variables. Methods We performed an experimental phantom study with 4 mediastinal iodine phantoms. Each phantom contained 6 tubes of different iodine concentrations (0, 1.0, 2.5, 5.0, 10.0, and 20.0 mg/mL) diluted in a specific solvent, which was water, 10% amino acid solution, 20% lipid emulsion, or 18% calcium solution, respectively. Mediastinal phantoms were inserted into an anthropomorphic chest phantom and were scanned with 3 different DECT scanners from 3 vendors using 2 radiation dosage settings. Iodine density (IoD) and computed tomography (CT) attenuation at virtual monoenergetic 70-keV images and virtual nonenhanced images were measured for the iodine phantoms. The effects of DECT scanners, solvents, and radiation dosage on the absolute measurement error of IoD and on the CT attenuation profiles were investigated using linear mixed-effects models. Measurement variability range of IoD was also determined. Results Absolute error of IoD was not significantly affected by the DECT systems, kind of solvents, and radiation dosage settings. However, CT attenuation profiles were significantly different among the DECT vendors and simulated body fluids. Measurement variability range of IoD was from -0.6 to 0.4 mg/mL for the true iodine concentration 0 mg/mL. Conclusions Dual-energy CT systems and fluid characteristics did not have a significant effect on the IoD measurement accuracy. A cutoff of IoD for the determination of a truly enhancing lesion on DECT would be 0.4 mg/mL.

Original languageEnglish
Pages (from-to)321-327
Number of pages7
JournalInvestigative Radiology
Volume53
Issue number6
DOIs
StatePublished - 2018.06.1

Keywords

  • dual-energy computed tomography
  • iodine density
  • iodine quantification
  • material decomposition
  • measurement variability