Skip to main navigation Skip to search Skip to main content

Design of millimeter-sized coils for power transmission to in vivo robotic capsules

  • J. H. Son*
  • , J. S. Hwang
  • , K. M. Song
  • , M. H. Ryu
  • , J. D. Kim
  • , S. Baang
  • , S. W. Lee
  • , J. W. Kim
  • *Corresponding author for this work
  • Hallym University
  • Biomedlab Co., Ltd.

Research output: Contribution to conferenceConference paperpeer-review

Abstract

This paper suggests an arrangement method of coils to provide hundreds of mW power via inductive link to in vivo robotic capsules, whose orientation are practically undetermined. We propose that the orthogonal coils in 3 dimensions could provide power amounting from sixty to one hundred percent of the maximum possible power of one receiving coil regardless of the orientations of the machines.

Original languageEnglish
Title of host publicationProceedings of the 3rd IASTED International Conference on Biomedical Engineering 2005
EditorsM.H. Hamza
Pages499-502
Number of pages4
StatePublished - 2005
Event3rd IASTED International Conference on Medical Engineering 2005 - Innsbruck, Austria
Duration: 2005.02.162005.02.18

Publication series

NameProceedings of the 3rd IASTED International Conference on Biomedical Engineering 2005

Conference

Conference3rd IASTED International Conference on Medical Engineering 2005
Country/TerritoryAustria
CityInnsbruck
Period05.02.1605.02.18

Keywords

  • And in vivo robotic capsules
  • Coil design
  • Inductive link
  • Receiving coil
  • Wireless power transmission

Fingerprint

Dive into the research topics of 'Design of millimeter-sized coils for power transmission to in vivo robotic capsules'. Together they form a unique fingerprint.

Cite this