Abstract
A sensitive and selective method for the determination of 4'-ethyl-3-methyl-3-piperidinopropiophenone hydrochloride (eperisone hydrochloride) in human plasma was developed and validated. The procedure employed an internal standard and a solvent extraction step followed by chromatography on a Xterra C18 minibore column. Detection was by electrospray ionization tandem mass spectrometry with multiple reaction monitoring. The mass transitions of eperisone and tolperisone (IS) were m/z 260 → 98 and m/z 246 → 98, respectively. The method has a limit of detection of 0.1 pg/mL for eperisone based on the three times signal-tonoise value with a linear range from 0.01 to 10.0 ng/mL for the analyte. Extraction recovery was on average 98.6±7.2% (SD) for eperisone. The Intra- and inter-day assay accuracy ranged from 93 to 114% and precision (RSD) was better than 8.5%. The method was successfully employed to analyze plasma samples and evaluate pharmacokinetics of eperisone in healthy male volunteers.
| Original language | English |
|---|---|
| Pages (from-to) | 1174-1188 |
| Number of pages | 15 |
| Journal | Archives of Pharmacal Research |
| Volume | 30 |
| Issue number | 9 |
| DOIs | |
| State | Published - 2007.09.30 |
Keywords
- Eperisone
- Human plasma
- Pharmacokinetics
- Tandem mass spectrometry
- Tolperisone
Fingerprint
Dive into the research topics of 'Determination of eperisone in human plasma by liquid chromatography-ESI- tandem mass spectrometry'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver