Abstract
In the present study, a Compton camera, including detailed timing characteristics of the component detectors and related data acquisition (DAQ) electronics, was modeled very accurately with Geant4. The results are very convincing - the inclusion of the timing characteristics in the simulation reduced the imaging sensitivity discrepancy between the simulated and measured data to a negligible level (12%) for a 22Na point source; this is a significant accuracy improvement compared with the results obtained previously without modeling the timing characteristics, for which the discrepancies were 30% and 6% for 511 and 1275 keV, respectively. The imaging resolution accuracy was also significantly improved.
| Original language | English |
|---|---|
| Pages (from-to) | S274-S275 |
| Journal | Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment |
| Volume | 633 |
| Issue number | SUPPL. 1 |
| DOIs | |
| State | Published - 2011.05 |
Keywords
- Compton camera
- Geant4
- Monte Carlo
- Timing characteristics