Skip to main navigation Skip to search Skip to main content

Intensity modulated radiosurgery for the spine: Dosimetric impact of beamlet size variation in the leaf travel direction

  • Joo Young Kim
  • , Hyun Soo Shin
  • , Ja Young Kim
  • , Hyeli Park
  • , Sung Joon Kim
  • CHA University

Research output: Contribution to journal › Journal article › peer-review

Abstract

Background and purpose: To investigate the dosimetric impact of beamlet size in the leaf travel direction for the spinal treatment using intensity-modulated radiosurgery (IMRS). Materials and methods: The IMRS plans of ten patients (11 lesions - 6 thoracic, 2 cervical, 3 lumbar) were re-planned using four different beamlet sizes (1, 2, 5, and 10 mm) - in the leaf travel direction, while keeping the Y-dimension by multi-leaf collimator (MLC) width fixed, and compared to the reference plan with beamlet size of 3 mm. To evaluate the beamlet size effect, target volumes (coverage, conformity, and size effect), organ at risks (OARS) (doses to the spinal cord, lung and kidneys), and integral dose, and monitor units (MUs) were calculated. Results: Target coverage and dose conformity for planning target volume (PTV) were not correlated with beamlet size. Maximum (p = 0.000) and mean (p = 0.000) spinal cord doses decreased by 4.0% and 3.4% from 23.4% and 28.6% as beamlet size decreased from 10 to 1 mm. The integral doses, MUs and doses to other organs increased at smaller beamlet sizes. MUs for a beamlet size of 10 mm decreased by 31.4%, as compared with that at the reference beamlet size. Conclusions: Despite no dosimetric benefits with respect to target volume and an MU increase, a definite dose reduction was observed at the spinal cord for smaller beamlet sizes. Treatment with IMRS planning for the spine will benefit from the use of a beamlet size between 2 and 4 mm.

Original languageEnglish
Pages (from-to)227-233
Number of pages7
JournalRadiotherapy and Oncology
Volume98
Issue number2
DOIs
StatePublished - 2011.02
Externally publishedYes

Keywords

  • Beamlet
  • Dose conformity
  • Intensity-modulated radiosurgery (IMRS)
  • Spinal cord
  • Target coverage

Fingerprint

Dive into the research topics of 'Intensity modulated radiosurgery for the spine: Dosimetric impact of beamlet size variation in the leaf travel direction'. Together they form a unique fingerprint.

Cite this