Not all green is tophi: The importance of optimizing minimum attenuation and using a tin filter to minimize clumpy artifacts on foot and ankle dual-energy CT

  • Pusan National University
  • Jeonbuk National University
  • Siemens Healthineers Ltd

Research output: Contribution to journalJournal articlepeer-review

Abstract

OBJECTIVE: The clumpy artifact has a high misdiagnosis rate, but the artifact has not been well studied. The aims of this study were to evaluate the frequency and location of clumpy artifacts, the rate of misdiagnosis of clumpy artifacts as gout, and the effects of raising the minimum attenuation value and using a selective photon shield in dual-energy CT (DECT). MATERIALS AND METHODS: Forty patients without gout who underwent foot and ankle DECT were enrolled in this study. Images in both sets were randomly assigned a minimum attenuation of 130 HU or 150 HU. Three radiologists independently checked all images for presence, volume, and location of green color-coded pixelation and graded their findings according to a 4-point confidence scale, frequency, and volume. Misdiagnosis rate and misdiagnosis score were compared using the Wilcoxon signed rank and McNemar tests. RESULTS: In set 1, the frequency of clumpy artifacts in DECT with the minimum attenuation set to 130 HU and 150 HU were 81% and 68%, respectively. For all three readers, the misdiagnosis rate and misdiagnosis score decreased when changing the minimum attenuation from 130 HU to 150 HU. In set 2, with the minimum attenuation set to 130 HU, the frequency of the clumpy artifact was 44%; with the minimum attenuation set to 150 HU, no clumpy artifacts were seen. CONCLUSION: Clumpy artifacts occurred frequently in DECT without a tin filter. Setting the minimum attenuation to the higher value of 150 HU reduced the frequency of clumpy artifacts, and adding a tin filter to DECT greatly reduced their occurrence.

Original languageEnglish
Pages (from-to)1335-1342
Number of pages8
JournalAmerican Journal of Roentgenology
Volume214
Issue number6
DOIs
StatePublished - 2020.06

Keywords

  • Clumpy artifact
  • Dual-energy CT
  • Gout
  • Pseudotophi

Quacquarelli Symonds(QS) Subject Topics

  • Medicine

Fingerprint

Dive into the research topics of 'Not all green is tophi: The importance of optimizing minimum attenuation and using a tin filter to minimize clumpy artifacts on foot and ankle dual-energy CT'. Together they form a unique fingerprint.

Cite this