Optimization of DPC dSiPM-Based DOI Compton Camera by Monte Carlo Simulation

  • Jae Hyeon Kim
  • , Young Su Kim
  • , Hyun Su Lee
  • , Hee Seo
  • , Chan Hyeong Kim
  • Hanyang University
  • Korea Atomic Energy Research Institute

Research output: Contribution to journalJournal articlepeer-review

Abstract

In this paper, a Compton camera capable of estimating depth of interaction (DOI) in the detector was designed and optimized using Geant4. The proposed DOI Compton camera consists of two pixelated CsI(Tl) scintillators coupled with digital silicon photomultipliers [Digital Photon Counter (DPC)-3200-22]. The parameters affecting the performance of a Compton camera were evaluated for optimization in terms of image sensitivity and image resolution with a measure of figure of merit. The considered parameters were: 1) type of the reflector surface; 2) thickness of the scintillator; and 3) the distance between the scatter and the absorber detectors. The simulation results showed that a ground white surface is better than a polished white one. The optimal thickness and the scatter-absorber distance were found to be 15 mm and 4 cm, respectively. With the optimized parameters, performance of the DOI Compton camera was estimated for a 137Cs source at 1 m using Geant4. The image resolutions were 6.06° full-width at half-maximum with the list-mode maximum likelihood expectation maximization algorithm. It took 20 s to take Compton images for the gamma ray sources of 0.1 μ Sv/h intensity with the DOI Compton camera. It was shown that there was a clear improvement by employing DOI estimation algorithm in combination with the use of thick scintillators for a Compton camera.

Original languageEnglish
Pages (from-to)1424-1431
Number of pages8
JournalIEEE Transactions on Nuclear Science
Volume65
Issue number7
DOIs
StatePublished - 2018.07

Keywords

  • Compton imaging
  • depth of interaction (DOI)
  • digital silicon photomultiplier (dSiPM)
  • Monte Carlo
  • optimization study
  • radiation measurement