Abstract
Photo Acoustic Tomography (PAT) is a hybrid imaging modality which combines high contrast of optical imaging and spatial resolution of ultrasound imaging, thus it is suitable to image biological tissue noninvasively. Laser-induced photoacoustic signals were measured from a sample by means of an unfocused ultrasound transducer, then PAT image was reconstructed based on a universal back-projection algorithm. To evaluate the feasibility of our system, phantom test was performed, consequently, the PAT images obtained using our system showed highly analogous shape and volume with those of the phantom. This result demonstrated that our system can provide a powerful tool for imaging the substructure of biological tissue in non-invasive manner.
| Original language | English |
|---|---|
| Pages (from-to) | 2349-2351 |
| Number of pages | 3 |
| Journal | Transactions of the Korean Institute of Electrical Engineers |
| Volume | 59 |
| Issue number | 12 |
| State | Published - 2010.12 |
Keywords
- Back-projection algorithm
- Photoacoustic tomography
- Unfocused ultrasound transducer
Fingerprint
Dive into the research topics of 'Phantom evaluation and development of photoacoustic tomography imaging system using unfocused ultrasound transducer and back-projection algorithm'. Together they form a unique fingerprint.Cite this