Prevention of platelet adhesion on polysulfone porous catheter by saline solution perfusion, II. Ex vivo and in vivo investigation

  • Gilson Khang
  • , Joon B. Park*
  • , Hai Bang Lee
  • *Corresponding author for this work
  • University of Iowa
  • Korea Research Institute of Chemical Technology

Research output: Contribution to journalJournal articlepeer-review

Abstract

Using a custom-made PVC medical tube ateriovenous (A-V) shunt and a porous polysulfone (PSf) catheter, an ex vivo and in vivo experiment were carried out to study the effectiveness of saline solution perfusion to prevent adhesion of platelets without using anticoagulants. The critical perfusion rate for preventing adhesion of platelets onto the PSf catheter tubes is about 0.45 ml min-1 cm-2. The higher perfusion rate resulted in a lesser platelet adhesion. At the highest perfusion rate tested, 0.92 ml min-1 cm-2, the number of adhered platelets decreased by 98% in the in vivo experiment. From the results of ex vivo and in vivo canine experiments, the saline perfusion catheters can be used as a very useful clinical armamentarium to prevent platelet and blood cell adhesion without using anticoagulants.

Original languageEnglish
Pages (from-to)123-134
Number of pages12
JournalBio-Medical Materials and Engineering
Volume6
Issue number2
DOIs
StatePublished - 1996

Keywords

  • Blood compatibility
  • Ex vivo and in vivo canine experiment
  • Platelet adhesion
  • Polysulfone porous catheter tube
  • Saline perfusion