Quantitative computed tomographic evaluation of bone mineral density in beagle dogs: Comparison with dual-energy x-ray absorptiometry as a gold standard

  • Danbee Kwon
  • , Jaehwan Kim
  • , Hyeeun Lee
  • , Byeol Kim
  • , Hyunhee Han
  • , Hyunju Oh
  • , Minjung Kim
  • , Hakyoung Yoon
  • , Byeongchun Lee
  • , Kidong Eom*
  • *Corresponding author for this work
  • Konkuk University
  • Seoul National University

Research output: Contribution to journalJournal articlepeer-review

Abstract

This study aimed to demonstrate the higher accuracy and reproducibility of quantitative computed tomography (QCT) compared with dual-energy X-ray absorptiometry (DXA) as a gold standard for measuring canine bone mineral density (BMD). Seven middle-aged beagle dogs underwent lumbar vertebral and bilateral femoral DXA and QCT scans. BMD (mg/cm2) was measured at the vertebral body from L2 to L6, femoral neck, and proximal and distal femoral diaphyses. The BMD values were measured 3 times and compared. The BMD value on QCT was higher than that on DXA for femoral BMD but not for vertebral BMD. The correlation was strong for the lumbar vertebrae (r=0.66) and was strongest for L3 (r=0.85). No correlation was found for the femoral neck (P=0.35), and only moderate correlations were found for the proximal and distal femoral diaphyses (r=0.43 and r=0.40, respectively). The limits of agreement were narrower for vertebral BMD than for femoral BMD, and L3 had the narrowest limits of agreement. The intraclass correlation (ICC) was higher for DXA than for QCT at all lumbar and femoral sites measured, but the ICC of QCT was higher than 0.7. In conclusion, L3 can be used to monitor changes in BMD, and relative values and sequential monitoring of femoral BMD can also be useful because of the high reproducibility of QCT measurements. QCT would be a useful technique for evaluation of BMD in veterinary practice.

Original languageEnglish
Pages (from-to)620-628
Number of pages9
JournalJournal of Veterinary Medical Science
Volume80
Issue number4
DOIs
StatePublished - 2018.04

Keywords

  • Bone mineral density (BMD)
  • Dog
  • Dual-energy X-ray absorptiometry (DXA)
  • Quantitative computed tomography (QCT)

Fingerprint

Dive into the research topics of 'Quantitative computed tomographic evaluation of bone mineral density in beagle dogs: Comparison with dual-energy x-ray absorptiometry as a gold standard'. Together they form a unique fingerprint.

Cite this