Skip to main navigation Skip to search Skip to main content

Validation of automatically measured monophasic action potential recordings

  • Jong Min Lee
  • , Gi Byoung Nam
  • , Kyoungsuk Rhee
  • , Kee Joon Choi
  • , Sun I. Kim
  • , You Ho Kim*
  • *Corresponding author for this work
  • Hanyang University
  • University of Ulsan

Research output: Contribution to journalJournal articlepeer-review

Abstract

This study devised an automatic monophasic action potential (MAP) measurement program, and compared the computer-measured MAP durations with those measured manually by two independent observers in order to facilitate the analysis of MAP data obtained during a clinical electrophysiology study. The results were compared at various cycle lengths and during pharmacologic or physiologic interventions. This program identified the onset, plateau, and baseline accurately using the MAP data. The automatically measured MAP durations at the 90% repolarization level (MAPd) strongly correlated with those measured manually (r=0.99). The MAPd shortened in parallel as the pacing cycle length was progressively shortened from 800 to 600, 400 and 300 ms.

Original languageEnglish
Pages (from-to)577-589
Number of pages13
JournalComputers in Biology and Medicine
Volume34
Issue number7
DOIs
StatePublished - 2004.10

Keywords

  • Automatic measurement
  • Monophasic action potential
  • Validation

Fingerprint

Dive into the research topics of 'Validation of automatically measured monophasic action potential recordings'. Together they form a unique fingerprint.