Suitable analytical procedures for the bulk analysis of ultra-trace amounts of uranium and plutonium have been developed using multi-collector inductively coupled mass spectrometry (MC–ICP–MS). The quantification and determination of the isotopic ratios of uranium and plutonium in three simulated swipe samples, a swipe blank, and a process blank were performed to validate the analytical performance. The analytical results for the simulated swipe samples were in good agreement with the certified values, based on the measurement quality goals for the analysis of bulk environmental samples recommended by the International Atomic Energy Agency (IAEA).